User Tools

Site Tools


cardiac_mri:protocols:structural

This is an old revision of the document!


Structural

<callout type=“info” icon=“true”> Identical locations must be used for each sequence if possible. Always used the “Copy image location” tool when setting up subsequent sequences </callout>

  1. Localizer
    • 3 plane
    • Repeat in ref if needed to move heart to isocenter, then repeat in ref and delete all but final loc
  2. Pseudo-SA
    • Perpindicular to long axis Loc
  3. 4ch CINE
    • 3PT at pseudo-SA RV free wall, LV free wall, apex from loc
    • Avoid the aortic valve
  4. 2ch CINE
    • 3PT at pseudo-SA LV anterior wall, LV inferior wall, from previous CINE
  5. 3ch CINE
    • 3PT at pseudo-SA aortic valve, pseudo-SA opposite point to AoV, apex from previous CINE
  6. LVOT CINE
    • Perpendicular to 3ch CINE
  7. Aortic Valve CINE
    • Slice thickness 5 mm
    • Peripindicular to LVOT through the AoV
  8. SA CINE Stack
    • Perpindicular to 4 chamber. Stack at 10 mm intervals to cover the entire left ventricle
  9. Axial HASTE
    • Free breathing, include the entire aortic arch to diaphragm
  10. Axial SSFP
    • Free breathing, include the entire aortic arch to diaphragm
  11. Aortic Valve Phase Contrast
    • Copy image location from AoV CINE
cardiac_mri/protocols/structural.1694727115.txt.gz · Last modified: by 127.0.0.1